The RC-110's top channel is an overflow bin: every event above the last calibrated energy is counted there, so it is not spectral content at that energy. It shows up as a single-channel spike around a hundred times its neighbours - 186 counts of 91224 in the indoor background capture, and 618 of 81225 (0.76% of all counts, in one channel) in a new 2.5 h forest background. No template can reproduce a one-channel spike, so leaving it in the Poisson likelihood does not invent a nuclide - it hides them. The only way the fit can cover the spike is by lifting the topmost continuum bump, and that raised continuum then absorbs genuine counts across the whole high-energy region. Measured, toggling FIT_OVERFLOW_CHANNELS 0/1 over the real captures: capture residual reduced chi2 overflow bin background 6.8 -> 6.4% 2.18 -> 1.69 0.20% of counts forest 9.7 -> 7.2% 3.72 -> 1.28 0.76% of counts thorium 11.7 -> 11.5% 1.94 -> 1.84 - uranium 8.7 -> 8.5% 2.01 -> 1.84 - Goodness-of-fit improves on all four and by far the most on the capture with the largest artefact, which is what ties the effect to this cause. Verdicts and time-to-identification are unchanged: background 900 s, thorium 120 s, uranium 300 s, forest 3600 s. Significance moves modestly in both directions - K-40 rises 3.7 -> 4.6 sigma on the forest capture but falls 6.8 -> 6.2 on the indoor one. The change is justified by the artefact being an artefact and by the fit becoming well-specified (chi2 1.3-1.8 where it was 1.9-3.7), not by larger numbers. The post-fix figures are the trustworthy ones precisely because the model now fits. The peak search already skipped this channel; the template fit, which is what actually decides identification now, did not. Also adds: - host/spectra/forest.csv - a real 9002 s / 81225 count outdoor background from a forest, 9.0 cps against the indoor capture's 21.7, now a permanent regression case for the low-rate end. - tools/xml_to_replay_csv.py - converts a RadiaCode XML export into the replay harness format. tools/radiacode_to_background.py writes counts/second references for the device's background library, which replay cannot consume, so there was no way to get an XML export under test. Co-Authored-By: Claude Opus 5 <noreply@anthropic.com>
98 lines
3.2 KiB
Python
98 lines
3.2 KiB
Python
#!/usr/bin/env python3
|
|
"""Convert a RadiaCode XML export into the replay-harness CSV format.
|
|
|
|
The harness reads what SdStorage::exportSpectrum() writes: duration and
|
|
calibration in header comments, then absolute counts per channel. The existing
|
|
tools/radiacode_to_background.py writes counts/SECOND references for the
|
|
device's background library instead, which replay cannot consume.
|
|
|
|
python xml2replay.py in.xml out.csv [--zero-last] [--dose D]
|
|
|
|
--zero-last clears the final channel, which on the RC-110 is an overflow bin
|
|
holding every event above the last calibrated energy rather than real spectral
|
|
content at that energy.
|
|
"""
|
|
import re
|
|
import sys
|
|
import xml.etree.ElementTree as ET
|
|
|
|
|
|
def strip_ns(tag):
|
|
return tag.split('}', 1)[-1]
|
|
|
|
|
|
def find_first(root, name):
|
|
for el in root.iter():
|
|
if strip_ns(el.tag) == name:
|
|
return el
|
|
return None
|
|
|
|
|
|
def main():
|
|
args = [a for a in sys.argv[1:]]
|
|
zero_last = '--zero-last' in args
|
|
if zero_last:
|
|
args.remove('--zero-last')
|
|
dose = None
|
|
if '--dose' in args:
|
|
i = args.index('--dose')
|
|
dose = float(args[i + 1])
|
|
del args[i:i + 2]
|
|
src, dst = args[0], args[1]
|
|
|
|
root = ET.parse(src).getroot()
|
|
spec = find_first(root, 'Spectrum')
|
|
counts = [int(float(dp.text)) for dp in spec
|
|
if strip_ns(dp.tag) == 'DataPoint']
|
|
|
|
cal = find_first(root, 'EnergyCalibration')
|
|
coeffs = [float(c.text) for c in cal.iter()
|
|
if strip_ns(c.tag) == 'Coefficient']
|
|
while len(coeffs) < 3:
|
|
coeffs.append(0.0)
|
|
a0, a1, a2 = coeffs[:3]
|
|
|
|
mt = find_first(root, 'MeasurementTime')
|
|
duration = int(float(mt.text))
|
|
|
|
if len(counts) != 1024:
|
|
sys.exit('expected 1024 channels, got %d' % len(counts))
|
|
|
|
overflow = counts[-1]
|
|
if zero_last:
|
|
counts[-1] = 0
|
|
|
|
total = sum(counts)
|
|
cps = total / duration if duration else 0.0
|
|
# The RC-110's published sensitivity is ~77 cps per uSv/h for Cs-137; only
|
|
# used when the caller does not supply a measured dose.
|
|
if dose is None:
|
|
dose = cps / 77.0
|
|
|
|
with open(dst, 'w', newline='\n') as f:
|
|
f.write('# RadiaCode-110 spectrum export (from %s)\n' % src.replace('\\', '/').split('/')[-1])
|
|
f.write('# duration_s,%d\n' % duration)
|
|
f.write('# total_counts,%d\n' % total)
|
|
f.write('# cps,%.3f\n' % cps)
|
|
f.write('# dose_uSv_h,%.4f\n' % dose)
|
|
f.write('# calib_a0,%.6f\n' % a0)
|
|
f.write('# calib_a1,%.6f\n' % a1)
|
|
f.write('# calib_a2,%.8f\n' % a2)
|
|
f.write('# background_subtracted,0\n')
|
|
f.write('channel,energy_keV,counts\n')
|
|
for ch, c in enumerate(counts):
|
|
e = a0 + a1 * ch + a2 * ch * ch
|
|
f.write('%d,%.2f,%d\n' % (ch, e, c))
|
|
|
|
print('%s -> %s' % (src, dst))
|
|
print(' %d ch, %d s, %d counts, %.2f cps, dose %.4f uSv/h' %
|
|
(len(counts), duration, total, cps, dose))
|
|
print(' calib a0=%.4f a1=%.5f a2=%.7f' % (a0, a1, a2))
|
|
print(' last-channel overflow bin held %d counts (%.2f%% of total)%s' %
|
|
(overflow, 100.0 * overflow / max(total + (overflow if zero_last else 0), 1),
|
|
' - ZEROED' if zero_last else ''))
|
|
|
|
|
|
if __name__ == '__main__':
|
|
main()
|